Configuring Series Settings
Introduction
You can customize the following settings for your specific site and study in the Series settings tab:
Path
Settings > Image Display > Series
Steps
- Modify the following settings as needed.
- When you have finished configuring all settings, click Close to apply and save setting changes.
MR Series Settings
Use the MR Series settings to configure the DICOM Echo and Temporal Position Indicator (TPI) series as a separate or a combined series.
Setting |
Description |
---|---|
Separate MR series by echo number |
If enabled, Synapse separates images of the same series into multiple series groups using the DICOM Echo tag. This setting is enabled by default. |
Separate MR series by temporal position indicator |
If enabled, Synapse separates images of the same series into multiple series groups using the DICOM TPI tag. This setting is enabled by default. |
Separate MR series by image continuity |
If enabled, Synapse separates an MR series into separate image sequences to ensure the sequences have uniform spacing and direction. Note: For enhanced MR mulitframes, Synapse does not use the Multi-frame Dimension Module to split and sort images. This setting is enabled by default. |
CT and PT Series Settings
Use the CT and PT Series settings to configure how Synapse displays CT, PT, and localizer images. You can split CT or PT series, containing localizer and non-localizer images, into two separate series. You can also separate a CT or PT series when the images are non-sequential, have gaps, have the same position, or have different orientations.
Note: |
|
Setting |
Description |
---|---|
Separate CT series by localizer |
If enabled, Synapse separates localizer and non-localizer images from the same series into two series groups. This setting is enabled by default. |
Separate CT or PT series by image continuity |
If enabled, Synapse separates a CT or PT series into separate image sequences to ensure the sequences have uniform spacing and direction. Note: For enhanced CT multiframe, Synapse does not use the Multi-frame Dimension Module to split and sort images. This setting is enabled by default. |
Display localizer series in reading protocol
|
If enabled, Synapse displays localizer images in the Viewer for modality and procedure-based reading protocols. This setting is enabled by default. |
CT localizer series position
|
Select one of the following options:
Synapse displays the localizer series first in the Series Picker and in the modality-match reading protocols. |
MG Series Settings
Use the MG Series settings to configure how Synapse displays mammography images. Synapse also supports MG Tomosynthesis series.
Setting |
Description |
---|---|
Zoom to breast bounds on initial display |
If enabled, Synapse:
or
Note: Synapse might not set an image to Zoom to Breast Bounds if the image compression is set to lossy. This setting is enabled by default. |
Show CAD |
If enabled, Synapse automatically shows Computer-Aided Detection (CAD) markers when you open a study. Note: This option is only available only for MG studies. This setting is disabled by default. |
Propagate implant suppression to all images |
If enabled, Synapse does the following:
If disabled, you can toggle breast implant suppression for the active viewport by clicking the Implant Suppression On/Off icon. Note:
This setting is disabled by default. |
Implant suppression on initial display |
If enabled, Synapse applies breast implant suppression for 2D and tomosynthesis mammography studies by reducing the breast implant brightness on initial display. Note:
This setting is disabled by default. |
PET Series Settings
Use the PET Series settings to determine how Synapse displays the initial top and bottom values for an SUV PET series. These settings affect the brightness, contrast, and appearance of the image for viewing, as you prefer. You can select a value that includes a maximum of four decimal places. The settings you select will persist and be available on any workstation you log into.
Setting |
Description |
---|---|
Enable Initial Display |
Select the check box to enable the setting and then adjust to the Top and Bottom settings, described in the following two rows of this table. This setting is disabled by default. |
Top |
Click plus (+) or minus (-) for each value to select a Top value that is greater than the Bottom value. The default Top setting is: 3 |
Bottom |
Click plus (+) or minus (-) for each value to select a Bottom value. The default Bottom setting is: 0 |
Series Order Options Settings
Use the Series Order Options settings to configure how Synapse sorts and displays a series:
Setting |
Description |
---|---|
Order the All images series according to the Series Picker thumbnail order |
If enabled, Synapse sorts the All Images series by Series Picker order. If disabled, Synapse sorts the All Images series according to Synapse rules order. |
Split XA/US single-frame images into separate thumbnails |
If enabled, Synapse splits XA/US single frame images into separate thumbnails. If disabled, Synapse combines single frame images into one thumbnail, which displays as the first thumbnail in Series Picker. Images within the single frame series are sorted oldest to newest. This setting is disabled by default. |
Series Picker Display Settings
Use the Series Picker Display settings to configure how Synapse displays the Series Picker and the Anywhere Series Picker. Your System Administrator or a user with the appropriate privilege can enable or disable these settings at the Enterprise, Role, or User level using Administrator Mode.
Setting |
Description |
---|---|
Series Picker split scroll bar
|
Select this option to split the Series Picker toolbar into two toolbars, one for the current study and one for prior studies. This setting is disabled by default. |
Disable Thumbnail Series Description |
Select one or more modalities to exclude the thumbnail series description in the Series Picker toolbar and the Anywhere Series Picker. This setting is disabled by default for all modalities. |
Localizer Handling Settings
Use the Localizer Handling settings to configure how Synapse displays localizer series. For CT, PT, and MR studies, the Display localizer series in reading protocol setting is enabled by default and will display the localizer series at the beginning or at the end of a study in both the Viewer and the Series Picker based on the Localizer series position setting status.
Note: |
For MR studies, Synapse uses a heuristic algorithm to identify a localizer series, if possible. |
---|
Setting |
Description |
---|---|
Display localizer series in reading protocol |
If enabled, Synapse displays localizer images in the Viewer for modality-match reading protocols. Note: This setting applies to all reading protocol types. This setting is enabled by default. |
Localizer series position |
Select one of the following options:
Using the default setting, Synapse displays the localizer series first in the Series Picker and in the modality-match reading protocols. When you select the Last setting, Synapse displays in the Series Picker after all series containing DICOM images, but before the All Images and Bookmark series. |
Results
Synapse automatically saves any updated options and displays the following message in the bottom left corner of the window: Successfully saved.
Synapse displays the following message to alert you that an action must be taken for the settings to apply: Refresh the Synapse browser window for settings to apply.